Please use this identifier to cite or link to this item:
http://localhost:8080/xmlui/handle/123456789/5477
Full metadata record
DC Field | Value | Language |
---|---|---|
dc.contributor.author | R Flaih, Khalid | - |
dc.date.accessioned | 2022-10-23T08:35:18Z | - |
dc.date.available | 2022-10-23T08:35:18Z | - |
dc.date.issued | 2020-08-24 | - |
dc.identifier.uri | http://localhost:8080/xmlui/handle/123456789/5477 | - |
dc.description.abstract | "In this work, the dependence of the bulk etches rate VB for CR-39detector on the temperature T, etching time t and concentration N of the NaOH etchant solution has been studied for detectors material types PADC. The concentration, etching time, temperature and other conditions have been varied in the ranges (5.25-8.25) N, (1-5) hand (60-80) °C for PADC respectively. In this case, the thickness of material removed by etching ranges from 0.20 to 6.10𝜇m/hr for CR-39detector.The 6.25 N NaOH solution at 70°C with etching time 3 hours represents the optimum etching condition for nuclear detector CR-39 " | en_US |
dc.publisher | Journal of University of Anbar for Pure Science (JUAPS) | en_US |
dc.subject | "Nuclear detectors CR-39 ; bulk etch rate VB; Chemical etchant NaOH " | en_US |
dc.title | Parameters Affecting Bulk Etch Rate VB for CR-39 Detector | en_US |
Appears in Collections: | قسم الفيزياء |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
Parameters Affecting - khalid flyah.pdf | 498.93 kB | Adobe PDF | View/Open |