Please use this identifier to cite or link to this item: http://localhost:8080/xmlui/handle/123456789/7577
Title: Determination of Aniline Blue dye by Flow Injection Analysis With Home Made Valve
Authors: SALEH FARHOOD, Ahmed
AHMED MOHAMMED ALI, Luma
Fadhel Ali, Firas
Keywords: Merging zone
Valve
Aniline blue
Dead volume
Dispersion
Reproducibility
Issue Date: 2017
Publisher: Oriental Journal of Chemistry
Abstract: Flow injection unit with merging zone technique was designed for aniline blue determination spectrophotometrically. This unit consists of a valve was designed for dye injection. The scope of the study is to find the optimum conditions :volume of sample, flow rate of carrier, dead volume, dispersion coefficient and reproducibility. They found 117µl, 4.3 ml/min, zero dead volume, 2.62 and high reproducibility for six times respectively. The range of aniline blue concentration by flow injection analysis was 0. 1 – 8 ppm and detection limit 0.1 with R2 = 0.9952 and 0.01 – 16 ppm for spectrophotometric method with R2 =0.9947 with detection limit 0.01 .
URI: http://localhost:8080/xmlui/handle/123456789/7577
ISSN: 0970-020 X
Appears in Collections:قسم الكيمياء

Files in This Item:
File Description SizeFormat 
Determination of Aniline Blue - Firas Ali.pdf452.29 kBAdobe PDFView/Open